Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131
Check against the official notice below — it lists the full affected range.
Fort Defiance Industries is recalling all Automated Field Steam Sterilizers due to a defect in the water recovery system's 120V Amphenol connector. The connector may experience intermittent loss of ground continuity, which could create an unreliable fault current return path. The recall affects units distributed worldwide, including those in the United States (Maryland, Virginia, New Jersey, and Tennessee) and internationally in Canada, Lithuania, and Spain.
The recall applies to all serial numbers of this device. However, the recall record does not specify a remedy or instructions for what affected consumers should do.
Written by software from the official notice below. Not reviewed by a lawyer.
The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.
Reason for recall: The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path. Distribution: Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain. Lot/code info: Lot Code: All serial numbers/ GTIN: (01) 00862992000327