← Search recalls
FDA (Device enforcement)Recalled 2026-07-08class ii

Tobii Dynavox Llc: TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798

Is my product affected?

TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798

  • UPC 240628000004
  • UPC 260617070001

Check against the official notice below — it lists the full affected range.

What happened

Tobii Dynavox is recalling the TD Navio Midi Speech Generating Device due to a potential safety hazard affecting certain units. The battery mounting tabs in affected devices may become damaged under certain conditions, which could allow the battery to move inside the device. This movement could potentially cause an internal short circuit, leading to localized heating and gas venting within the device enclosure.

The recall affects devices distributed nationwide in the US with serial numbers TDNM1-240628000004 through TDNM1-260617070001. Devices with serial numbers starting from TDNM1-260617070001 and onward (production batch 20260615) have an improved design and are not part of this recall. The source document does not specify what remedy consumers should pursue.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.

Details from the source

Reason for recall: Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure. Distribution: US Nationwide distribution. Lot/code info: UDI-DI 073400747602056 Serial Numbers TDNM1-240628000004 through TDNM1-260617070001. Corrected/future production is differentiated by new serialization from TDNM1-260617070001 onward (production batch 20260615, improved design).

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.