TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798
Check against the official notice below — it lists the full affected range.
Tobii Dynavox is recalling the TD Navio Midi Speech Generating Device due to a potential safety hazard affecting certain units. The battery mounting tabs in affected devices may become damaged under certain conditions, which could allow the battery to move inside the device. This movement could potentially cause an internal short circuit, leading to localized heating and gas venting within the device enclosure.
The recall affects devices distributed nationwide in the US with serial numbers TDNM1-240628000004 through TDNM1-260617070001. Devices with serial numbers starting from TDNM1-260617070001 and onward (production batch 20260615) have an improved design and are not part of this recall. The source document does not specify what remedy consumers should pursue.
Written by software from the official notice below. Not reviewed by a lawyer.
Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.
Reason for recall: Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure. Distribution: US Nationwide distribution. Lot/code info: UDI-DI 073400747602056 Serial Numbers TDNM1-240628000004 through TDNM1-260617070001. Corrected/future production is differentiated by new serialization from TDNM1-260617070001 onward (production batch 20260615, improved design).
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.