TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799
Check against the official notice below — it lists the full affected range.
Tobii Dynavox LLC is recalling TD Navio Maxi Speech Generating Devices due to a potential battery safety hazard. Under certain conditions, the battery mounting tabs in the device may become damaged, causing the battery to shift inside the device enclosure. This movement could lead to an internal short circuit, resulting in localized heating and gas venting from within the device.
The recall affects TD Navio Maxi devices distributed nationwide in the United States with serial numbers TDNL1-240904000016 through TDNL1-260617070001. Devices with serial numbers TDNL1-260617070001 and later incorporate an improved design and are not part of this recall. The source document does not specify what remedy affected consumers should pursue.
Written by software from the official notice below. Not reviewed by a lawyer.
Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.
Reason for recall: Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure. Distribution: US Nationwide distribution. Lot/code info: UDI-DI 073400747602070 Serial Numbers TDNL1-240904000016 through TDNL1-260617070001. Corrected/future production is differentiated by new serialization from TDNL1-260617070001 onward (production batch 20260615, improved design).
Official record: FDA (Device enforcement) ↗