← Search recalls
FDA (Device enforcement)Recalled 2018-04-06closedclass ii

Orthofix, Inc: H-Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) 12MM Spacer (003-H-12001); b) 15MM Spacer (003-H-15001); 20MM Spacer (003-H-20001); 25MM Spacer (003-H-

Is my product affected?

H-Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) 12MM Spacer (003-H-12001); b) 15MM Spacer (003-H-15001); 20MM Spacer (003-H-20001); 25MM Spacer (003-H-25001); 30 MM (003-H-30001)

  • UPC 18056099647908
  • UPC 18056099647915
  • UPC 18056099647939
  • UPC 18056099647953
  • UPC 18056099647960

Check against the official notice below — it lists the full affected range.

What happened

Orthofix, Inc. is voluntarily recalling the Rival Reduce Fracture Plating System and Rival View Plating System due to a defect in which the locking screws do not properly engage the plates during surgery. The recalled products are H-Plate spacers in 12MM, 15MM, 20MM, and 25MM sizes manufactured on November 13, 2017. A total of 75 units were distributed across multiple U.S. states including New Jersey, Florida, Pennsylvania, South Carolina, Louisiana, Virginia, Minnesota, Texas, Georgia, and others.

The recall affects specific lot codes and UDI numbers identified in the record. The source document does not specify what actions affected consumers should take.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.

Details from the source

Reason for recall: Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery. Distribution: U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK, Quantity: 75 Lot/code info: Lot: B1212400, Date of Manufacture: 2017-11-13, Use By: 2022-10-13, (UDI)/Lot: a) (18056099647908) B1096740; b) (18056099647915) B1096733; c) (18056099647939) B1096734; d) (18056099647953) B1096735; e) (18056099647960) B1100226

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.