← Search recalls
FDA (Device enforcement)Recalled 2018-04-06closedclass ii

Orthofix, Inc: Locking Screw, D 2.0MM X 8MM, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label. Size (Part Number): a) D 2.0MM X 8MM (002-E-20008); b) D 2.0MM X 10MM (002-E-20010); c) D 2.0MM X 12MM (0

Is my product affected?

Locking Screw, D 2.0MM X 8MM, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label. Size (Part Number): a) D 2.0MM X 8MM (002-E-20008); b) D 2.0MM X 10MM (002-E-20010); c) D 2.0MM X 12MM (002-E-20012); d) D 2.0MM X 14MM (002-E-20014); e) D 2.0MM X 16MM (002-E-20016); f) D 2.0MM X 18MM (002-E-20018); g)

  • UPC 18056099646161
  • UPC 18056099646178
  • UPC 18056099646185
  • UPC 18056099646192
  • UPC 18056099646208
  • UPC 18056099646215
  • UPC 18056099646222
  • UPC 18056099646239
  • UPC 18056099646246
  • UPC 18056099646253
  • UPC 18056099646260
  • UPC 18056099646277

Check against the official notice below — it lists the full affected range.

What happened

Orthofix, Inc. is voluntarily recalling locking screws used in the Rival Reduce Fracture Plating System and Rival View Plating System. During a complaint investigation, the company confirmed that these screws do not properly engage the plates during surgery. The affected screws come in three sizes (2.0MM x 8MM, 2.0MM x 10MM, and 2.0MM x 12MM) and were distributed to medical facilities across 20 states and Washington, D.C. A total of 870 units were affected, spanning multiple lot codes.

The recall includes specific lot and UDI codes for the affected products. The source document does not specify a remedy or recommended action for affected consumers.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.

Details from the source

Reason for recall: Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery. Distribution: U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK, Quantity: 870 Lot/code info: (UDI)/Lot: a) (18056099646161)B1204974, B1089899, B1201171, B1201171; b) (18056099646178) B1195696, B1204268, B1083839; c) (18056099646185) B1215486, B1201172, B1201172, B1204281, B1083845; d) (18056099646192) B1215489, B1204975, B1083855, B1201173, B1201173; e) (18056099646208) B1215482, B1204585, B1195748, B1089887; f) (18056099646215) B1215488, B1204559, B1083856; g) (18056099646222) B1215483, B1204558, B1083854; h) (18056099646239) B1215490, B1204277, B1195721, B1083849; i) (18056099646246) B1215484, B1204279, B1195703, B1089890; j) (18056099646253) B1215491, B1204283, B1195720, B1083648; k) (18056099646260) B1215485, B1089879, B1201174, B1204267, B1204267; l) (18056099646277) B1215487, B1204284, B1195751, B1083857

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify