Rival Reduce Kit Plates, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Part Number: 000-K-012C
Check against the official notice below — it lists the full affected range.
Orthofix, Inc. is voluntarily recalling the Rival Reduce Fracture Plating System and Rival View Plating System (Part Number 000-K-012C) due to a confirmed defect in which locking screws do not properly engage the plates during surgery. The affected product is a sterile, prescription-only medical device manufactured by Orthofix S.r.l. in Italy, with 50 units distributed across multiple U.S. states including California, Texas, Florida, and others.
The recall affects devices with UDI 18056099648608 and Lot B1107421. The source document does not specify a remedy or instructions for what affected consumers should do.
Written by software from the official notice below. Not reviewed by a lawyer.
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Reason for recall: Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery. Distribution: U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK, Quantity: 50 Lot/code info: UDI: 18056099648608, Lot: B1107421,