TRIDENT 0¿ POLY INSERT 28mm ID 620-00-28D Orthopaedic implant components manufactured from UHMWPE (polyethylene)
Check against the official notice below — it lists the full affected range.
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Reason for recall: Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation. Distribution: Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa. Quantity: 12,156 US Lot/code info: 620-00-28D UDI-DI 07613327313581 Lots MHANP2 MHD1N0