← Search recalls
FDA (Device enforcement)Recalled 2020-06-18closedclass ii

Philips North America, LLC: Phillips Charging Station, Product #: 989803191021 - Product Usage: This charging station is used in the Expression MR400 MRI Patient Monitoring System (Model MR400) is int

Is my product affected?

Phillips Charging Station, Product #: 989803191021 - Product Usage: This charging station is used in the Expression MR400 MRI Patient Monitoring System (Model MR400) is intended to monitor vital signs for patients undergoing MRI procedures and to provide signals for synchronization for the MRI scann

  • UPC 989803191021

Check against the official notice below — it lists the full affected range.

What happened

Philips North America, LLC is recalling 118 Philips Charging Stations (Product #989803191021) used with the Expression MR400 MRI Patient Monitoring System. The charging stations contain Philips LI-Polymer rechargeable batteries that may not charge properly when inserted into a charger bay. When affected, the LED indicator light displays solid red and the battery will not charge, which may result from poor electrical contact between the battery and the charger.

The recalled units were manufactured prior to March 2020 and were distributed worldwide. The source document does not specify a remedy or action for affected consumers.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

A problem has been detected with the Philips LI-Polymer rechargeable battery when used with its approved charging station and module charger. When the battery is inserted into a charger bay, the LED indicatory lights solid red and the battery will not charge. This may result from poor electrical contact between the battery and charger.

Details from the source

Reason for recall: A problem has been detected with the Philips LI-Polymer rechargeable battery when used with its approved charging station and module charger. When the battery is inserted into a charger bay, the LED indicatory lights solid red and the battery will not charge. This may result from poor electrical contact between the battery and charger. Distribution: World wide distribution. Firm reports distributing more product then quantity manufactured. Clarification requested. And received. Firm made error in counting number of batteries distributed. Quantity: 118 units Lot/code info: Product #: 989803191021 Manufacturing date codes: prior to 2020-03

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.