Phillips 3.7V Rechargeable Li-Polymer Battery, Product #: 989803191341 - Product Usage: This battery is used in the Expression MR400 MRI Patient Monitoring System (Model MR400) is intended to monitor vital signs for patients undergoing MRI procedures and to provide signals for synchronization for th
Check against the official notice below — it lists the full affected range.
Philips North America, LLC has recalled 24,545 rechargeable 3.7V Li-Polymer batteries (Product #989803191341) used in Expression MR400 MRI Patient Monitoring Systems. The batteries were manufactured between December 2018 and February 2020 and were distributed worldwide. The recalled batteries experience a problem when inserted into their approved charging station or module charger: the LED indicator light turns solid red and the battery fails to charge, which may result from poor electrical contact between the battery and charger.
The recall record does not specify a remedy or corrective action for affected consumers.
Written by software from the official notice below. Not reviewed by a lawyer.
A problem has been detected with the Philips LI-Polymer rechargeable battery when used with its approved charging station and module charger. When the battery is inserted into a charger bay, the LED indicatory lights solid red and the battery will not charge. This may result from poor electrical contact between the battery and charger.
Reason for recall: A problem has been detected with the Philips LI-Polymer rechargeable battery when used with its approved charging station and module charger. When the battery is inserted into a charger bay, the LED indicatory lights solid red and the battery will not charge. This may result from poor electrical contact between the battery and charger. Distribution: World wide distribution. Firm reports distributing more product then quantity manufactured. Clarification requested. And received. Firm made error in counting number of batteries distributed. Quantity: 24,545 Units Lot/code info: Product #: 989803191341 Manufacturing date codes 2018-12-DD to 2020-02-DD
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.