← Search recalls
FDA (Device enforcement)Recalled 2020-05-21closedclass ii

Elekta Limited: Elekta Unity, Philips Marlin Software versions: R5.3.30, R5.3.31, R6.0.5331, R5.7.0, R5.7.1, R6.1.571

Is my product affected?

Elekta Unity, Philips Marlin Software versions: R5.3.30, R5.3.31, R6.0.5331, R5.7.0, R5.7.1, R6.1.571

  • UPC 05060191071321

Check against the official notice below — it lists the full affected range.

What happened

Elekta Limited is recalling Elekta Unity and Philips Marlin radiation therapy software (versions R5.3.30, R5.3.31, R6.0.5331, R5.7.0, R5.7.1, and R6.1.571) distributed in the United States and internationally. The software has a defect in its Treatment Session Manager Motion Monitoring feature that can cause a misalignment between the overlay of organ contours and the actual motion monitoring images displayed during treatment for certain anatomical areas including the abdomen, head and neck, pelvis, and thorax. A total of 26 units were affected by this issue.

The recall notice does not specify what actions affected consumers or facilities should take in response to this recall.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

In TSM (Treatment Session Manager) Motion Monitoring workflows, under certain conditions there can be a mismatch between the contour data overlay with respect to the motion monitoring images of the monitored structure.

Details from the source

Reason for recall: In TSM (Treatment Session Manager) Motion Monitoring workflows, under certain conditions there can be a mismatch between the contour data overlay with respect to the motion monitoring images of the monitored structure. Distribution: US distribution to IA, NY, PA, TX, WI. International distribution to Australia, Canada, China, Denmark, Germany, Hong Kong, Italy, Netherlands, Sweden, Switzerland, Turkey, United Kingdom. Quantity: 26 units Lot/code info: Software versions: R5.3.30, R5.3.31, R6.0.5331, R5.7.0, R5.7.1, R6.1.571. UDI Number 05060191071321. Anatomy (Protocol Name): Abdomen (btFFE Cor RealTime, btFFE Cor Sag RealTime, btFFE Cor Sag Tra RealTime, btFFE Fast Sag Cor RealTime, btFFE Sag RealTime); HeadNeck L (bFFE Cor RealTime, bFFE Cor Sag RealTime, bFFE Cor Sag Tra RealTime, bFFE Sag RealTime, bFFE Sag Tra RealTime); HeadNeck M (bFFE Cor RealTime, bFFE Cor Sag RealTime, bFFE Cor Sag Tra RealTime, bFFE Sag RealTime, bFFE Sag Tra RealTime); Pelvis L (bFFE Cor RealTime, bFFE Cor Sag RealTime, bFFE Cor Sag Tra RealTime, bFFE Sag RealTime, bFFE Sag Tra RealTime); Pelvis M (bFFE Cor RealTime, bFFE Cor Sag RealTime, bFFE Cor Sag Tra RealTime, bFFE Sag RealTime, bFFE Sag Tra RealTime); Thorax (btFFE Cor RealTime, btFFE Cor Sag RealTime, btFFE Cor Sag Tra RealTime, btFFE Sag RealTime, btFFE Sag RealTime)

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify