RayStation 6, UDI # 07350002010013 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
Check against the official notice below — it lists the full affected range.
RayStation 6, a software system used for radiation therapy and medical oncology treatment planning, has been recalled due to a software error found in its Pencil Beam Scanning and Line Scanning dose calculation algorithms. This error could cause the software to underestimate radiation doses in certain areas. The affected software has UDI # 07350002010013 with build number 6.0.0.24 and was distributed nationwide across multiple states including Florida, Georgia, Utah, Virginia, California, Maryland, Arizona, Texas, Washington DC, Missouri, Massachusetts, Illinois, Oklahoma, New Jersey, Tennessee, and Louisiana.
The recall record does not specify a remedy or recommended action for affected consumers.
Written by software from the official notice below. Not reviewed by a lawyer.
Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.
Reason for recall: Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose. Distribution: US Nationwide distribution including in the states of Florida, Georgia, Utah, Virginia, California, Maryland, Arizona, Texas, Washington DC, Missouri, Massachusetts, Illinois, Oklahoma, New Jersey, Tennessee, and Louisiana. Quantity: N/A Lot/code info: UDI # 07350002010013 Build number: 6.0.0.24
Official record: FDA (Device enforcement) ↗