FDA (Device enforcement)Recalled 2014-11-20closedclass ii

Roche Diagnostics Operations, Inc.: ACCU-CHEK Tender I 17/110 10 pieces Product Usage: Intended for the infusion and/or injection of fluids into the body below the surface of the skin. The indwelling

Is my product affected?

ACCU-CHEK Tender I 17/110 10 pieces Product Usage: Intended for the infusion and/or injection of fluids into the body below the surface of the skin. The indwelling catheter can be inserted independently from the infusion catheter and can be accessed for the injections through the injection port. The

  • UPC 04541421001

Check against the official notice below — it lists the full affected range.

What happened

Roche Diagnostics Operations, Inc. is recalling ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets (catalog number 04541421001, lots 5040427, 5044352, and 5061933) due to a potential defect where the tubing can become detached at the connect/disconnect location. These are devices intended for infusing or injecting fluids under the skin. A total of 13,041 boxes were distributed nationwide in the US.

If the tubing detaches during use, insulin delivery will stop without the pump alarming to alert the user to the interruption. The recall notice does not specify a remedy for affected consumers.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

Roche Diabetes Care was informed by UnoMedical, the Manufacturer of ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets, about potential for the the tubing of the infusion set to become detached at the connect/disconnect location on the ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets. If tubing detachment occurs, insulin delivery is interrupted and the pump will not alarm to notify

Details from the source

Reason for recall: Roche Diabetes Care was informed by UnoMedical, the Manufacturer of ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets, about potential for the the tubing of the infusion set to become detached at the connect/disconnect location on the ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets. If tubing detachment occurs, insulin delivery is interrupted and the pump will not alarm to notify Distribution: US Nationwide Distribution Quantity: Total of 13,041 boxes Lot/code info: Catalog: 04541421001 and lots: 5040427, 5044352, and 5061933.

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify