ACCU-CHEK Tender I 17/60 10 pieces Product Usage: Intended for the infusion and/or injection of fluids into the body below the surface of the skin. The indwelling catheter can be inserted independently from the infusion catheter and can be accessed for the injections through the injection port. The
Check against the official notice below — it lists the full affected range.
Roche Diagnostics Operations, Inc. is recalling ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets due to a potential defect where the tubing can become detached at the connect/disconnect location. These infusion sets are used to deliver insulin and other fluids under the skin. If the tubing detaches, insulin delivery stops, and importantly, the pump will not sound an alarm to alert the user to this problem. A total of 13,041 boxes were distributed nationwide.
The affected products have catalog number 04541430001 and lot numbers 5040425, 5044367, 5051305, 5053294, and 5061932. The source document does not specify what consumers should do regarding this recall.
Written by software from the official notice below. Not reviewed by a lawyer.
Roche Diabetes Care was informed by UnoMedical, the Manufacturer of ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets, about potential for the the tubing of the infusion set to become detached at the connect/disconnect location on the ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets. If tubing detachment occurs, insulin delivery is interrupted and the pump will not alarm to notify
Reason for recall: Roche Diabetes Care was informed by UnoMedical, the Manufacturer of ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets, about potential for the the tubing of the infusion set to become detached at the connect/disconnect location on the ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets. If tubing detachment occurs, insulin delivery is interrupted and the pump will not alarm to notify Distribution: US Nationwide Distribution Quantity: Total of 13,041 boxes Lot/code info: Catalog: 04541430001 and lots: 5040425, 5044367, 5051305, 5053294, and 5061932.