Identified an issue with Selenia Dimensions 1.12.0 and 3Dimensions 2.3.0 system software when used in combination with C-View software in which the quality of the final C-View synthesized 2D image is impacted and may appear blurry and result in misdiagnosis with the possibility being a false negative diagnosis
Reason for recall: Identified an issue with Selenia Dimensions 1.12.0 and 3Dimensions 2.3.0 system software when used in combination with C-View software in which the quality of the final C-View synthesized 2D image is impacted and may appear blurry and result in misdiagnosis with the possibility being a false negative diagnosis Distribution: Worldwide distribution - US Nationwide and the countries of Belgium, Canada, Israel, Korea, Republic of, Netherlands. Quantity: 425 units Lot/code info: UDI-DI: 15420045505636, 15420045511897 Selenia Dimensions 1.12.0 and 3Dimensions 2.3.0 system software when used in combination with C-View¿ software
15420045505636, 15420045511897
Official record: FDA (Device enforcement) ↗