ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
Reason for recall: ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met. Distribution: US Nationwide distribution in the states of MO, MN, MA, OH, NE. Quantity: 360 units Lot/code info: UDI-DI: 00191072198402; Lot: 2405152; Exp: 05/15/2025
00191072198402