If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management
Reason for recall: If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management Distribution: US: MS, WA, TX, NJ, GA, PN, NY, AL, MN, NE, IN, CT, CA, IL, TN, WI, ND, NM, AR, NC, PR, RI, WY, MD, FL, HI, AZ, SC, MO, OK, VA, ME, KS, MI, LA, IA, OH, SD, NH, KY, AL, OR, MA, GU. OUS: Afghanistan, Albania, Algeria, Andorra, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia-Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Cameroon, Canada, Ch Quantity: 7022 Lot/code info: 11335791, UDI-DI: 04056869152059: 1.0/VA10G and prior, 1.5/ VB10H and prior , 2.0/ VB11C and prior, 2.5/ VB30D and prior. 11653093, UDI-DI: 04056869957357: 2.0/ VB11C and prior, 2.5/ VB30D and prior.
04056869152059, 04056869957357