← Search recalls
FDA (Device enforcement)Recalled 2024-09-03class ii

Philips North America Llc: Ingenia Ambition X - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extr

Hazard

Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.

Details from the source

Reason for recall: Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement. Distribution: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Iceland, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Maldives, Mauritius, Mexico, Morocco, Netherlands, Pakistan, Peru, Poland, Portugal, Qatar, Romania, Russian Quantity: 63 units Lot/code info: UDI-DI: (01)00884838098350(21) Serial Numbers; 47558 47602 47606 47608 47609 47614 47615 47617 47621 47622 47625 47626 47628 47630 47634 47644 47646 47651 47654 47656 47657 47660 47665 47666 47667 47668 47671 47672 47674 47677 47678 47679 47680 47681 47684 47685 47687 47688 47694 47695 47697 47698 47700 47701 47702 47703 47705 47706 47708 47709 47710 47716 47733 47737 47738 47753 47755 47759 47785 47791 47800 47802 47804

UPC codes

00884838098350

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.