Atellica CH Microalbumin_2 (¿ALB_2) and the ADVIA Chemistry Microalbumin_2 (¿ALB_2) are not meeting the High-Dose Hook Effect claim/Prozone Effect claim as stated in the Instructions for Use (IFU) on the Atellica CH, Atellica CI Analyzers, ADVIA 1800 Chemistry Systems, ADVIA 2400 Chemistry Systems, and ADVIA Chemistry XPT Systems, may result in erroneously depressed microalbumin patient results.
Reason for recall: Atellica CH Microalbumin_2 (¿ALB_2) and the ADVIA Chemistry Microalbumin_2 (¿ALB_2) are not meeting the High-Dose Hook Effect claim/Prozone Effect claim as stated in the Instructions for Use (IFU) on the Atellica CH, Atellica CI Analyzers, ADVIA 1800 Chemistry Systems, ADVIA 2400 Chemistry Systems, and ADVIA Chemistry XPT Systems, may result in erroneously depressed microalbumin patient results. Distribution: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bosnia Herzeg, Brazil, Canada, Chile, Colombia, Croatia, Cura¿ao, St. Eus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hong¿Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latv Quantity: 18483 units Lot/code info: UDI-DI: 00630414596310 Lot Numbers: 232033, 232128, 232137, 232146, 232147, 242149, 242150, 242365
00630414596310