← Search recalls
FDA (Device enforcement)Recalled 2026-08-12class i

Abiomed, Inc.: Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Contr

Is my product affected?

Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, IN; Product Code: 0042-0000-IN. 4

  • UPC 00813502010022
  • UPC 00813502011289
  • UPC 00813502011418
  • UPC 00813502010428
  • UPC 00813502011296

Check against the official notice below — it lists the full affected range.

Hazard

The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.

Details from the source

Reason for recall: The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly. Distribution: Domestic: Nationwide Distribution. International: United Arab Emirates, Austria, Australia, Belgium, Brunei Darussalam, Brazil, Canada, Switzerland, Chile, China, Colombia, Costa Rica, Cyprus, Czechia, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Croatia, Hungary, Ireland, Israel, India, Italy, Japan, Kuwait, Luxembourg, Mexico, Malaysia, Netherlands, Norway, New Ze Quantity: 5232 units (3733 US, 1499 OUS) 36 3 Lot/code info: 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA; UDI-DI: 00813502010022; Serial Numbers: IC1324, IC1346, IC1480, IC1481, IC1543, IC1544, IC1572, IC1595, IC1598, IC1664, IC1665, IC1715, IC1769, IC1790, IC1791, IC1911, IC1912, IC1913, IC2032, IC2033, IC2034, IC2035, IC2369, IC2370, IC2686, IC2687, IC3661, IC3662, IC3802, IC3803, IC4944, IC4945, IC5034, IC5050, IC5680, IC5821, IC5824, IC6082, IC6083, IC6633, IC6634, IC6635, IC6636, IC7694, IC7695, IC7716, IC8283, IC9413, IC9414, IC9415, IC9416; 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU; UDI-DI: 00813502011289; Serial Numbers: IC1347, IC1349, IC6845, IC1285, IC1416, IC2765, IC2766, IC3021, IC3022, IC3718, IC3719, IC3720, IC3776, IC5558, IC5559, IC1578, IC1585, IC1902, IC1903, IC1905, IC2678, IC2679, IC2855, IC2856, IC3053, IC3054, IC3329, IC3671, IC3672, IC3830, IC3831, IC4033, IC4837, IC5086, IC5169, IC5171, IC5560, IC5681, IC5682, IC5683, IC6064, IC6081, IC6149, IC6270, IC6272, IC6437, IC6439, IC6441, IC6443, IC6810, IC6812, IC6813, IC6814, IC7261, IC7262, IC7393, IC7394, IC7576, IC7577, IC7646, IC1825, IC1826, IC1870, IC1876, IC4153, IC4175, IC4657, IC4658, IC5476, IC5477, IC5518, IC5520, IC7263, IC7264, IC1011, IC1490, IC1584, IC1590, IC1592, IC2175, IC2176, IC2857, IC2996, IC2997, IC3291, IC3296, IC3317, IC3318, IC3486, IC3488, IC3698, IC3702, IC3724, IC3778, IC3864, IC3870, IC3890, IC3891, IC4151, IC4152, IC4323, IC4324, IC4471, IC4501, IC4681, IC4682, IC4851, IC4892, IC4933, IC4934, IC

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify