During manufacturing, the last three tests from the 160 test flex of lot 23206BA (wells 8 or 10) were processed on the Dimension Vista platform. There was an issue with the filling line which led to some wells to underfill in lot 23206BA. Calibrator, quality control (QC) and patient results could have been impacted.
Reason for recall: During manufacturing, the last three tests from the 160 test flex of lot 23206BA (wells 8 or 10) were processed on the Dimension Vista platform. There was an issue with the filling line which led to some wells to underfill in lot 23206BA. Calibrator, quality control (QC) and patient results could have been impacted. Distribution: Domestic: Nationwide Distribution. International: Austria, Belgium, Canada, Czech Republic, Denmark, France, Germany, Italy, Qatar, Slovenia, Spain, Switzerland. Quantity: 5,371 units (4,394 US, 977 OUS) Lot/code info: Siemens Material Number (SMN): 10445146; UDI-DI: 00842768015717; Lot Number: 23206BA;
00842768015717