← Search recalls
FDA (Device enforcement)Recalled 2026-06-12class ii

Medline Industries, LP: Medline Convenience Kits: 1) BLOOD DRAW KIT, Catalog Number: DYLAB1081; 2) WHOLE BLOOD PHLEBOTOMY KIT, Catalog Number: DYNDB3005D; 3) QP KIT, Catalog Number: DYNDB3005F; 4) DOU

Is my product affected?

Medline Convenience Kits: 1) BLOOD DRAW KIT, Catalog Number: DYLAB1081; 2) WHOLE BLOOD PHLEBOTOMY KIT, Catalog Number: DYNDB3005D; 3) QP KIT, Catalog Number: DYNDB3005F; 4) DOUBLE RED CELL PHLEBOTOMY KIT, Catalog Number: DYNDB3006C; 5) PLATELET PHLEBOTOMY KIT, Catalog Number: DYNDB3007C; 6) WHOLE BL

  • UPC 10198459199608
  • UPC 40198459199609
  • UPC 10195327038700
  • UPC 40195327038701
  • UPC 10198459025068
  • UPC 40198459025069
  • UPC 10195327038717
  • UPC 40195327038718
  • UPC 10195327038724
  • UPC 40195327038725
  • UPC 10195327170738
  • UPC 40195327170739
  • UPC 10195327028565
  • UPC 40195327028566
  • UPC 10195327661236
  • UPC 40195327661237
  • UPC 10195327313234
  • UPC 40195327313235
  • UPC 10195327086275
  • UPC 40195327086276
  • UPC 10198459028915
  • UPC 40198459028916
  • UPC 10195327432300
  • UPC 40195327432301
  • UPC 10195327502492
  • UPC 40195327502493
  • UPC 10195327502614
  • UPC 40195327502615
  • UPC 10195327662462
  • UPC 40195327662463
  • UPC 10195327215835
  • UPC 40195327215836
  • UPC 10195327216047
  • UPC 40195327216048
  • UPC 10195327509873
  • UPC 40195327509874
  • UPC 10198459234965
  • UPC 40198459234966
  • UPC 10195327323332
  • UPC 40195327323333

Check against the official notice below — it lists the full affected range.

Hazard

The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the seal area. If the sterile barrier is compromised, patients and users may be exposed to microbial contamination. Potential health consequences may include no injury, localized infection, or, less commonly, more serious infection requiring medical treatment, antimicrobial therapy, procedural intervention, or hospitalization. In rare cases, particularly in critically ill or immunocompromised patients, severe infection could become life-threatening.

Details from the source

Reason for recall: The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the seal area. If the sterile barrier is compromised, patients and users may be exposed to microbial contamination. Potential health consequences may include no injury, localized infection, or, less commonly, more serious infection requiring medical treatment, antimicrobial therapy, procedural intervention, or hospitalization. In rare cases, particularly in critically ill or immunocompromised patients, severe infection could become life-threatening. Distribution: US Nationwide distribution. Quantity: 996692 kits Lot/code info: 1) DYLAB1081, UDI-DI: 10198459199608(each), 40198459199609(case), Lot Number: 25CMC514; 2) DYNDB3005D, UDI-DI: 10195327038700(each), 40195327038701(case), Lot Number: 24CBG614; 3) DYNDB3005D, UDI-DI: 10195327038700(each), 40195327038701(case), Lot Number: 24CBT912; 4) DYNDB3005D, UDI-DI: 10195327038700(each), 40195327038701(case), Lot Number: 23HBH298; 5) DYNDB3005D, UDI-DI: 10195327038700(each), 40195327038701(case), Lot Number: 23JBN521; 6) DYNDB3005D, UDI-DI: 10195327038700(each), 40195327038701(case), Lot Number: 23KBT270; 7) DYNDB3005F, UDI-DI: 10198459025068(each), 40198459025069(case), Lot Number: 25BBD241; 8) DYNDB3005F, UDI-DI: 10198459025068(each), 40198459025069(case), Lot Number: 25BBP477; 9) DYNDB3005F, UDI-DI: 10198459025068(each), 40198459025069(case), Lot Number: 25CBI161; 10) DYNDB3005F, UDI-DI: 10198459025068(each), 40198459025069(case), Lot Number: 24LBM458; 11) DYNDB3005F, UDI-DI: 10198459025068(each), 40198459025069(case), Lot Number: 24FBL961; 12) DYNDB3005F, UDI-DI: 10198459025068(each), 40198459025069(case), Lot Number: 24FBS200; 13) DYNDB3005F, UDI-DI: 10198459025068(each), 40198459025069(case), Lot Number: 24IBO183; 14) DYNDB3006C, UDI-DI: 10195327038717(each), 40195327038718(case), Lot Number: 24ABB637; 15) DYNDB3006C, UDI-DI: 10195327038717(each), 40195327038718(case), Lot Number: 24ABY532; 16) DYNDB3006C, UDI-DI: 10195327038717(each), 40195327038718(case), Lot Number: 24CBH932; 17) DYNDB3006C, UDI-DI: 10195327038717(each), 40195327038718(case), Lo

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify