← Search recalls
FDA (Device enforcement)Recalled 2026-06-12class ii

Medline Industries, LP: Medline Convenience Kits: 1) EYE KIT, Model Number: DYKM2751; 2) EYE KIT, Model Number: DYKM2751A; 3) CATARACT PREOP KIT, Model Number: DYKS1583A; 4) CATARACT PREOP KIT, Model

Is my product affected?

Medline Convenience Kits: 1) EYE KIT, Model Number: DYKM2751; 2) EYE KIT, Model Number: DYKM2751A; 3) CATARACT PREOP KIT, Model Number: DYKS1583A; 4) CATARACT PREOP KIT, Model Number: DYKS1583B

  • UPC 10198459118395
  • UPC 40198459118396
  • UPC 10198459313301
  • UPC 40198459313302
  • UPC 10198459146640
  • UPC 40198459146641
  • UPC 10198459292699
  • UPC 40198459292690

Check against the official notice below — it lists the full affected range.

Hazard

The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the seal area. If the sterile barrier is compromised, patients and users may be exposed to microbial contamination. Potential health consequences may include no injury, localized infection, or, less commonly, more serious infection requiring medical treatment, antimicrobial therapy, procedural intervention, or hospitalization. In rare cases, particularly in critically ill or immunocompromised patients, severe infection could become life-threatening.

Details from the source

Reason for recall: The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the seal area. If the sterile barrier is compromised, patients and users may be exposed to microbial contamination. Potential health consequences may include no injury, localized infection, or, less commonly, more serious infection requiring medical treatment, antimicrobial therapy, procedural intervention, or hospitalization. In rare cases, particularly in critically ill or immunocompromised patients, severe infection could become life-threatening. Distribution: US Nationwide distribution. Quantity: 2080 kits Lot/code info: 1) DYKM2751, UDI-DI: 10198459118395(each), 40198459118396(case), Lot Number: 24LBQ727; 2) DYKM2751, UDI-DI: 10198459118395(each), 40198459118396(case), Lot Number: 24JBB775; 3) DYKM2751A, UDI-DI: 10198459313301(each), 40198459313302(case), Lot Number: 25KMD727; 4) DYKS1583A, UDI-DI: 10198459146640(each), 40198459146641(case), Lot Number: 24KBP388; 5) DYKS1583A, UDI-DI: 10198459146640(each), 40198459146641(case), Lot Number: 24JBA739; 6) DYKS1583B, UDI-DI: 10198459292699(each), 40198459292690(case), Lot Number: 25DBM037

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify