Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.
Reason for recall: Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery. Distribution: US Nationwide distribution in the states of CA, FL, IL, NJ, OH, PA. Lot/code info: Lot Code: Model No NCOCO6010; UDI-DI ¿03760279380209; Lot Numbers 250700051, 251000044, and 251000045
03760279380209
Official record: FDA (Device enforcement) ↗