Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.
Reason for recall: Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery. Distribution: US Nationwide distribution in the states of CA, FL, IL, NJ, OH, PA. Lot/code info: Lot Code: Model No NCOCO1575; UDI-DI ¿03760279380087; Lot Number 250700007, 251000031, and 260300041
03760279380087
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.