Philips SureSigns VS3 Vital Signs Monitor, 863069 863070 863071 863072 863073 863074 The SureSigns VS3 monitor is for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare environments. Additionally, the monitor is intended for use
Check against the official notice below — it lists the full affected range.
Philips is recalling certain SureSigns VS3 Vital Signs Monitors because lithium ion batteries that have exceeded their expected lifespan can overheat or ignite. The batteries should be replaced every 3 years or after 300 charge-discharge cycles. While the monitors can display battery age and charge-discharge cycle information, the current labeling does not provide complete instructions on how to use this information to determine when replacement is necessary. An overheating battery may cause the device case to overheat, melt, or ignite, potentially injuring patients or nearby users or damaging property.
The recall affects approximately 207,439 Philips SureSigns VS3 Vital Signs Monitors distributed worldwide, including throughout the United States and Puerto Rico. The recall record does not specify a remedy or action that affected consumers should take.
Written by software from the official notice below. Not reviewed by a lawyer.
Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries should be replaced every 3 years or upon reaching 300 charge-discharge cycles. Although the Philips SureSigns VS & VM Monitors and View Station (VSV) can display actual information on either or both battery age and charge-discharge cycles, the existing labeling does not include full instructions on how to use this information to determine when to replace the battery. An overheated battery may in turn cause the device case to overheat and possibly melt or cause the device to ignite, which can cause injury to a patient, nearby users, or cause damage to property.
Reason for recall: Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries should be replaced every 3 years or upon reaching 300 charge-discharge cycles. Although the Philips SureSigns VS & VM Monitors and View Station (VSV) can display actual information on either or both battery age and charge-discharge cycles, the existing labeling does not include full instructions on how to use this information to determine when to replace the battery. An overheated battery may in turn cause the device case to overheat and possibly melt or cause the device to ignite, which can cause injury to a patient, nearby users, or cause damage to property. Distribution: Worldwide distribution: US (nationwide) including Puerto Rico and countries of: Afghanistan, Albania, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Brazil, Bulgaria, Canada, Cambodia, Chile, China, Colombia, Costa Rica, Cote D'Ivoire, Croatia, Czech Republic, Dominican Republic, El Salvador, Egypt, Estonia, Ethiopia, France, French Guiana, French Polynesia, Quantity: VM 96,324 Units; VS 110,424 Units; VSV 691 Units Lot/code info: UDI# VS3 (863069): 00884838000094, (863070): 00884838000100, (863071): 00884838000117, (863072): 00884838000124, (863073): 00884838000131, (863074): 00884838000148
Official record: FDA (Device enforcement) ↗