Innovasis, Inc: Anti-Torque, Vector, Part: LS-749 part of Vector Pedicle Screw System

Hazard

Pedicle screw system intended to provide immobilization/stabilization of spinal segments as adjunct to thoracic/lumbar/sacral spine fusion may have anti-torque instrument with undersized condition which could prevent mating tip from engaging with the pedicle screw tulip and rod during tightening, which may lead to surgical delay, intervention/alternative methods to perform final tightening step.

Details from the source

Reason for recall: Pedicle screw system intended to provide immobilization/stabilization of spinal segments as adjunct to thoracic/lumbar/sacral spine fusion may have anti-torque instrument with undersized condition which could prevent mating tip from engaging with the pedicle screw tulip and rod during tightening, which may lead to surgical delay, intervention/alternative methods to perform final tightening step. Distribution: US Nationwide distribution in the states of TX, CA. Quantity: 5 Lot/code info: LS-749: UDI-DI: M711BTA2020, M711LS7490, Serial Numbers: 002, 003, 007, 012, 014

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.