Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
Reason for recall: Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID. Distribution: US Nationwide distribution. Quantity: 21 Lot/code info: GEM Premier 5000: Part number: 00024019255 UDI: 08426950870227 Units distributed in Spain: GEM Premier 5000 Serial Nos.: 25028719, 18021546, 18021549, 26012036, 18051850, 23054764, 21068342, 25080219, 26010781, 25122012, 25041639, 25120558, 25079944, 25079943, 25028719, 25100485, 25079947, 25100483, 25100484, 21057870 and 26010710. Affected Accessory: Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026. Part No. 00024015859; UDI No. 08426950869634
00024019255, 08426950870227, 00024015859, 08426950869634