Affected kits contain recalled Bard Silastic Foley Catheters. Catheters were recalled by manufacturer due to increased amount of raw material present on the surface of the catheter. Issue may result in the use of urinary catheters that exhibit staining on the catheter surface; raw material present in increased amounts may be harmful.
Reason for recall: Affected kits contain recalled Bard Silastic Foley Catheters. Catheters were recalled by manufacturer due to increased amount of raw material present on the surface of the catheter. Issue may result in the use of urinary catheters that exhibit staining on the catheter surface; raw material present in increased amounts may be harmful. Distribution: US Nationwide distribution in the states of CA, FL, MD. PA, and TX. Quantity: 980 kits Lot/code info: DYNJ0945779U UDI-DI 10195327632656 Lots 25IMC004 25JMD942 25JMJ441 25LMG068
10195327632656
Official record: FDA (Device enforcement) ↗