← Search recalls
FDA (Device enforcement)Recalled 2026-06-11class ii

Boston Scientific Corporation: FARADRIVE STEERABLE SHEATH CLEAR, used to facilitate access to the vasculature and into the chambers of the heart, Material Number 21M402, IFU Reference Number LBL1062

Hazard

Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.

Details from the source

Reason for recall: Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body. Distribution: Global Distribution. US Nationwide. Quantity: 74,760 units Lot/code info: UDI-DI: 00810087180072; Lots: CL12697, CL12698, CL12701, CL12702, CL12824, CL12832, CL12833, CL12861, CL12888, CL12889, CL12912, CL12913, CL12916, CL12918, CL12968, CL12975, CL12977, CL13003, CL13016, CL13019, CL13045, CL13047, CL13059, CL13078, CL13079, CL13101, CL13102, CL13120, CL13121, CL13140, CL13141, CL13158, CL13159, CL13195, CL13211, CL13218, CL13232, CL13233, CL13259, CL13290, CL13339, CL13367, CL13400, CL13421, CL13516, CL13518, CL13572, CL13577, CL13615, CL13645, CL13669, CL13707, CL13720, CL13733, CL13821, CL13834, CL13857, CL13926, CL13941, CL13976, CL13989, CL14026, CL14052, CL14078, CL14121, CL14138, CL14170, CL14186, CL14225, CL14248, CL14285, CL14344, CL14352, CL14374, CL14395, CL14438, CL14487, CL14499, CL14517, CL14580, CL14632, CL14669, CL14690, CL14710, CL14748, CL14768, CL14967, CL15004, CL15008, CL15066, CL15106, CL15127, CL15190, CL15238, CL15271, CL15317, CL15357, CL15396, CL15426, CL15473, CL15489, CL15567, CL15636, CL15639, CL15753, CL15774, CL15807

UPC codes

00810087180072

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify