FDA (Device enforcement)Recalled 2024-07-24class ii

Stryker Corporation: Stryker iBur 3.0mm Coarse Diamond Round, Distal Bend - intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing

Hazard

Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.

Details from the source

Reason for recall: Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention. Distribution: Worldwide distribution - US Nationwide and the countries of Australia, Canada, France, Hong Kong, India, Italy, Japan, Japan, Netherlands, South Korea, Spain, Sweden, Taiwan, United Kingdom. Quantity: 2670 units Lot/code info: GTIN: 07613327501209 Lot Numbers: 22230017 22333017 22342017 23011017 23093017 23172017 23188017 23221017 23311017 24004027

UPC codes

07613327501209

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify