Some sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices.
Reason for recall: Some sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices. Distribution: Worldwide distribution - US Nationwide and the countries of Canada, Japan, EU, and Latin America. Quantity: 3,663 units Lot/code info: Model Number: 190A. UDI-DI: 00821925003538. All Lot Numbers.