Abbott Laboratories: Alinity hq Analyzer, REF: 09P68-01

Hazard

1)If Complete Blood Count with differential and reticulocyte(CBC+Diff+Retic) sample is run, then immediately following on same rack CBC+Diff sample is run, then falsely low red blood cell count may occur, generating falsely high Mean Cell Hemoglobin(MCH)/MCHC and falsely low hematocrit results. 2)Cell events may be incorrectly counted as basophil(BASO), resulting in increased BASO and %BASO counts

Details from the source

Reason for recall: 1)If Complete Blood Count with differential and reticulocyte(CBC+Diff+Retic) sample is run, then immediately following on same rack CBC+Diff sample is run, then falsely low red blood cell count may occur, generating falsely high Mean Cell Hemoglobin(MCH)/MCHC and falsely low hematocrit results. 2)Cell events may be incorrectly counted as basophil(BASO), resulting in increased BASO and %BASO counts Distribution: Worldwide - US Nationwide distribution in the states of NY, OK, NC and the countries of ALBANIA, ARGENTINA, AUSTRALIA, BANGLADESH, BELGIUM, BOSNIA AND HERZEGOVINA, Botswana, BRAZIL, BULGARIA, CANADA, CAYMAN ISLANDS, CHILE, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, FRANCE, GERMANY, GREECE, HONDURAS, HONG KONG, HUNGARY, INDIA, INDONESIA, ITALY, JAMAICA, JAPAN, JORDAN, KUWAIT, LEBANON, LITHU Quantity: 816 Lot/code info: UDI-DI: 380740138851. Software Versions: 5.7 (US) and 5.6 and below (OUS), Serial Numbers: HQ00105, HQ00108, HQ00117, HQ00118, HQ00119, HQ00120, HQ00126, HQ00127, HQ00128, HQ00132, HQ00136, HQ00139, HQ00140, HQ00141, HQ00142, HQ00144, HQ00145, HQ00147, HQ00148, HQ00149, HQ00152, HQ00153, HQ00154, HQ00156, HQ00157, HQ00161, HQ00162, HQ00166, HQ00169, HQ00171, HQ00172, HQ00173, HQ00174, HQ00175, HQ00183, HQ00184, HQ00185, HQ00186, HQ00187, HQ00188, HQ00266, HQ00268, HQ00269, HQ00270, HQ00271, HQ00273, HQ00274, HQ00276, HQ00277, HQ00278, HQ00279, HQ00281, HQ00282, HQ00283, HQ00286, HQ00287, HQ00289, HQ00291, HQ00295, HQ00296, HQ00299, HQ00300, HQ00302, HQ00303, HQ00312, HQ00313, HQ00314, HQ00316, HQ00318, HQ00319, HQ00321, HQ00324, HQ00327, HQ00328, HQ00329, HQ00331, HQ00332, HQ00334, HQ00335, HQ00336, HQ00337, HQ00338, HQ00417, HQ00418, HQ00419, HQ00420, HQ00422, HQ00424, HQ00425, HQ00430, HQ00432, HQ00433, HQ00434, HQ00435, HQ00436, HQ00438, HQ00439, HQ00440, HQ00441, HQ00442, HQ00443, HQ00444, HQ00445, HQ00446, HQ00447, HQ00448, HQ00451, HQ00453, HQ00455, HQ00456, HQ00457, HQ00458, HQ00460, HQ00461, HQ00463, HQ00464, HQ00466, HQ00468, HQ00469, HQ00471, HQ00472, HQ00474, HQ00475, HQ00476, HQ00536, HQ00539, HQ00540, HQ00541, HQ00542, HQ00544, HQ00545, HQ00546, HQ00547, HQ00548, HQ00550, HQ00551, HQ00552, HQ00553, HQ00554, HQ00555, HQ00558, HQ00561, HQ00562, HQ00563, HQ00564, HQ00565, HQ00567, HQ00568, HQ00569, HQ00570, HQ00571, HQ00573, HQ00574, HQ00575, HQ00576, HQ00577, HQ005

UPC codes

380740138851

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify