The weld on the light head frame can crack which could lead to problems such as the light head frame sagging, the plastic front face housing cracking, or the light head detaching.
Reason for recall: The weld on the light head frame can crack which could lead to problems such as the light head frame sagging, the plastic front face housing cracking, or the light head detaching. Distribution: Domestic: Nationwide Distribution International: Australia, Canada, Egypt, Hong Kong, Indonesia, Japan, Jordan, Netherlands, New Zealand, Philippines, Singapore, South America, South Korea, Thailand, UAE. Quantity: 822 units Lot/code info: Model Number: AUR7-LH/7TV-LH; Serial Numbers: 09010002, 9020003, 9020004, 9020007, 9020008, 9020009, 9020010, 9020011, 9020012, 9020013, 9020014, 9020015, 9020016, 9030017, 9030018, 9030019, 9030020, 9030021, 9030022, 9040023, 9040024, 9040025, 9040026, 9040027, 9040028, 9040029, 9040030, 9040031, 9040032, 9040033, 9040034, 9040035, 9040036, 9040037, 9040038, 9040039, 9040040, 9040041, 9040042, 9050043, 9050044, 9050045, 9050046, 9050047, 9050048, 9050049, 9050050, 9050051, 9050052, 9050053, 9050054, 9050055, 9050056, 9050057, 9050058, 9050059, 9050060, 9050061, 9050062, 9050063, 9050064, 9050065, 9050066, 9050067, 9050068, 9050069, 9050070, 9050071, 9050072, 9060073, 9060074, 9060075, 9060076, 9060077, 9060078, 9060079, 9060080, 9060081, 9060082, 9060083, 9060084, 9060085, 9060086, 9060087, 9060088, 9060089, 9060090, 9060091, 9060092, 9080093, 9080094, 9080095, 9080097, 9080098, 9080099, 9080100, 9080101, 9080102, 9080104, 9080105, 9080106, 9080107, 9090108, 9090109, 9090110, 9090111, 9090112, 9090113, 9090114, 9090115, 9090116, 9090117, 9090118, 9090119, 9110120, 9110121, 9110122, 9110123, 9110124, 9110125, 9110126, 9110127, 9110128, 9110129, 9110130, 9120131, 9120132, 9120133, 9120134, 9120135, 9120136, 9120137, 9120138, 9120139, 9120140, 9120141, 09120142, 9120143, 9120144, 10010145, 10010146, 10010147, 10010148, 10010149, 10010150, 10010151, 10010152, 10010153, 10010154, 10010155, 10010156, 10010157, 10010158, 10020159, 10020160, 10020161, 10020162, 10020163, 10020164, 1
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.