FDA (Device enforcement)Recalled 2017-05-12closedclass ii

DePuy Mitek, Inc., a Johnson & Johnson Co.: DePuy Mitek LATARJET EXPERIENCE Coracoid Top Hat Drill Product Code: 288202 Product Usage: The Latarjet Cortical Screw set is intended to treat recurrent sh

Is my product affected?

DePuy Mitek LATARJET EXPERIENCE Coracoid Top Hat Drill Product Code: 288202 Product Usage: The Latarjet Cortical Screw set is intended to treat recurrent shoulder instability by supporting the anteroinferior glenoid with a bony graft

  • UPC 10886705026807

Check against the official notice below — it lists the full affected range.

What happened

DePuy Mitek, Inc., a Johnson & Johnson company, has recalled 184 units of the Combo Screw Driver (Product Code 288211) used with the Latarjet Cortical Screw set, which is intended to treat recurrent shoulder issues. The screw driver tip has an increased potential to break during surgery when used at an angle off-axis to the screw. The affected products were distributed in the United States (Arizona, California, Colorado, Georgia, Massachusetts, Michigan, Ohio, and Washington) and internationally (Austria, Australia, Belgium, France, Germany, Spain, Switzerland, Netherlands, Poland, and the United Kingdom). The recalled lot codes are 16C01, 16E03, 16R01, 17A01, and 17B01.

The source document does not specify what action affected consumers or medical facilities should take regarding this recall.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

Combo Screw Driver (Product Code 288211) tip has the increased potential to break intra-operatively when being used at an angle off-axis to screw

Details from the source

Reason for recall: Combo Screw Driver (Product Code 288211) tip has the increased potential to break intra-operatively when being used at an angle off-axis to screw Distribution: US: AZ, CA, CO, GA, MA, MI, OH, WA Foreign: Austria, Australia, Belgium, France, Germany, Spain, Switzerland, Netherlands, Poland, UK Quantity: 184 units Lot/code info: GTN: 01)10886705026807 Lot codes: 16C01, 16E03, 16R01, 17A01, 17B01

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify