DePuy Mitek LATARJET EXPERIENCE Coracoid Top Hat Drill Product Code: 288202 Product Usage: The Latarjet Cortical Screw set is intended to treat recurrent shoulder instability by supporting the anteroinferior glenoid with a bony graft
Check against the official notice below — it lists the full affected range.
DePuy Mitek, Inc., a Johnson & Johnson company, has recalled 184 units of the Combo Screw Driver (Product Code 288211) used with the Latarjet Cortical Screw set, which is intended to treat recurrent shoulder issues. The screw driver tip has an increased potential to break during surgery when used at an angle off-axis to the screw. The affected products were distributed in the United States (Arizona, California, Colorado, Georgia, Massachusetts, Michigan, Ohio, and Washington) and internationally (Austria, Australia, Belgium, France, Germany, Spain, Switzerland, Netherlands, Poland, and the United Kingdom). The recalled lot codes are 16C01, 16E03, 16R01, 17A01, and 17B01.
The source document does not specify what action affected consumers or medical facilities should take regarding this recall.
Written by software from the official notice below. Not reviewed by a lawyer.
Combo Screw Driver (Product Code 288211) tip has the increased potential to break intra-operatively when being used at an angle off-axis to screw
Reason for recall: Combo Screw Driver (Product Code 288211) tip has the increased potential to break intra-operatively when being used at an angle off-axis to screw Distribution: US: AZ, CA, CO, GA, MA, MI, OH, WA Foreign: Austria, Australia, Belgium, France, Germany, Spain, Switzerland, Netherlands, Poland, UK Quantity: 184 units Lot/code info: GTN: 01)10886705026807 Lot codes: 16C01, 16E03, 16R01, 17A01, 17B01