← Search recalls
FDA (Device enforcement)Recalled 2026-05-20class i

Insulet Corporation: Omnipod 5 Pods. Includes the below Model/REF Numbers: 1. POD-OMNI-I1-6720. 2. POD-OMNI-I1-6220. 3. OMNI-I1-6729 (10-Pack), OMNI-I1-6720 (Single Pod). 4. OMNI-I1-6220. 5. OMNI-I1-6

Hazard

External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.

Details from the source

Reason for recall: External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death. Distribution: Worldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, Switzerland, Germany, Denmark, Finland, France, UK, Greece, Israel, Italy, Kuwait, Leichtenstein, Netherlands, Norway, Saudi Arabia, Sweden and Turkey. Quantity: 4,936,056 units Lot/code info: 1. Model/REF Number: POD-OMNI-I1-6720. Pack UDI-DI: 20385083000579. Single Pod UDI-DI: 10385083000527. Lot Numbers: PH1U01032521, PH1U01102521, PH1U12102421, PH1U12172421, PH1U11212421, PH1U11252421, PH1U11222421, PH1U12092421, PH1U09112421, PH1U11262421, PH1U04032521, PH1U12022421, PH1U12182421, PH1U04302524, PH1U05022521, PH1U01222521, PR1U11202533, PH1U09122421, PH1U10102421, PH1U12122421, PH1U01302522, PH1U02222521, PH1U05052521, PH1U10042421, PH1U12102412, PH1U01312521, PH1U03312521, PH1U05012521, PH1U09122422, PH1U09162421, PH1U09192421, PH1U09212421, PH1U09262411, PH1U09262421, PH1U10112422, PH1U10122421, PH1U10172421, PH1U12132422, PH1U01132521, PH1U01172521, PH1U02132511, PH1U03202521, PH1U03262521, PH1U04042521, PR1U11192531, PH1U11192511, PR1U01162621, PH1U09172421, PH1U09182421, PH1U09202421, PH1U10142421, PH1U10152421, PH1U10162421, PH1U01232522, PH1U01312511, PH1U02032521, PH1U02042521, PH1U02212521, PH1U03282521, PH1U04092521, PH1U04102521, PH1U04112521, PH1U10172522, PH1U10182521, PH1U10202521, PH1U10212521, PH1U10222521, PH1U10232521, PH1U10242521, PH1U10252521, PH1U11152521, PH1U11172511, PH1U11172521, PH1U11182521, PR1U11182531, PR1U11182541, PR1U11192541, PH1U11192521, PR1U11192532, PR1U11192542, PR1U11202521, PR1U11202532, PR1U11202541, PR1U11202542, PR1U11212521, PR1U11222531, PR1U11222541, PR1U11222521, PR1U11232541, PR1U11242521, PR1U11242531, PR1U11252541, PR1U11252512, PR1U11252521, PR1U11262521, PR1U12012521, PR1U12022521, PR1U12032521, PR1U12102522

UPC codes

20385083000579, 10385083000527, 20385083000425, 10385083000411, 2038508300043, 20385083000548, 20385083000135, 10385083000114

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify