The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
Reason for recall: The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement. Distribution: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Mauritius, Mexico, Netherlands, Norway, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., Quantity: 81 units Lot/code info: 1. Model Number: 728332; UDI-DI: 00884838059542; Serial Numbers: 860088, 860023, 860077, 860086, 60019, 860152, 860215, 60031, 60123, 60129, 60152, 860011, 860049, 860019, 860008, 860030, 860025, 860009, 860022, 860013, 860024, 860020, 860036, 860041, 860046, 860055, 860058, 860054, 860073, 860065, 860070, 860072, 860066, 860103, 860076, 860085, 860095, 860102, 860101, 860104, 860175, 860136, 860117, 860120, 860121, 860124, 860125, 860191, 860131, 860150, 860164, 60041, 860207, 60003, 860157, 860167, 860158, 860196, 60012, 860176, 860185, 860168, 860179, 860194, 60010, 60013, 60007, 60001, 395003, 860209, 860208, 60036, 60030, 60027, 60054, 60055, 60053, 60058, 60079, 395005, 60145.