← Search recalls
FDA (Device enforcement)Recalled 2023-04-25class ii

BioFire Diagnostics, LLC: FILMARRAY Gastrointestinal (GI) Panel (Part No.: RFIT-ASY-0104 & RFIT-ASY-0116) Pouch Module Software

Hazard

The firm has identified an increased risk of obtaining false positive Cryptosporidium results from customers using the BIOFIRE¿ FILMARRAY¿ Gastrointestinal (GI) Panel

Details from the source

Reason for recall: The firm has identified an increased risk of obtaining false positive Cryptosporidium results from customers using the BIOFIRE¿ FILMARRAY¿ Gastrointestinal (GI) Panel Distribution: US:DE, AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY. OUS: AD, AE, AI, AR, AT, AU, BD, BE, BF, BG, BH, BR, CA, CH, CL, CN, CO, CY, CZ, DE, DK, DZ, EE, ES, FI, FR, GA, GB, GE, GH, GM, GR, GU, HK, HU, IC, ID, IE, IL, IN, IQ, IT, JP, KE, KR, Quantity: 61,069 Lot/code info: Model/UDI-DI: RFIT-ASY-0104/00815381020116, RFIT-ASY-0116/00815381020109 using software versions prior to FLM2-SFW-0040 v2.0.3.1/BFS0001-5539 v2.0.200.10

UPC codes

00815381020116, 00815381020109

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify