Specimen collection device may leak after the patient sample swabs have been inserted into the test tubes, which could result in biohazard exposure, cross contamination causing false positive results for other specimens, and/or delay to test results.
Reason for recall: Specimen collection device may leak after the patient sample swabs have been inserted into the test tubes, which could result in biohazard exposure, cross contamination causing false positive results for other specimens, and/or delay to test results. Distribution: Worldwide distribution: US (nationwide): IL, TX, OH, LA, NY, CA, MD, IA, MN, KY, NH, WI, MO, NC, UT, WA, FL, NJ, SC, CO, PA, SD, NE, MT, WV, GA, MA, RI, CT, DE, MS, OK, VA, ID, AZ, OR, MI, IN, TN, AL, KS, ME, AR, NV, HI, AK, NM, WY, VT, ND, DC OUS (foreign) to countries of: Chile, Ireland, Oman, Nicaragua, Germany, San Marino, Ecuador, New Caledonia, France, Austria, United Kingdom, Finland, Italy Quantity: 651,150 Lot/code info: UDI-DI: 28053326001523. Lot/Expiration: 230397900/ July 3, 2024; 230535300/ July 14, 2024; 230627500/ August 4, 2024; 231877400/ April 10, 2025
28053326001523
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.