Hamilton Medical, Inc.: Hamilton-MR1, REF: 161010, Intensive Care Ventilator

Hazard

Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.

Details from the source

Reason for recall: Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation. Distribution: Distribution US nationwide, including Puerto Rico. OUS: Not provided Quantity: 12 units Lot/code info: UDI: 07630002800761 & 07630002813556 / Affected Serial Numbers: 3422 3423 3424 3425 3426 3427 3428 3429 3469 3503 3510 3800

UPC codes

07630002800761, 07630002813556

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.