← Search recalls
FDA (Device enforcement)Recalled 2026-04-22class ii

Intuitive Surgical, Inc.: Da Vinci ASSY, PSS, SP1098, Part Number: 380601

Hazard

Operative components of surgical systems potentially distributed with E-brake retainer pin, so in event of a structural column failure, the E-brake pin would prevent E-brake engagement and may result in uncontrolled motion. In addition, if it moves/backs out it may cause restricted column range of motion and/or uncontrolled motion, which may cause patient tissue injury/crushing injury/bleeding.

Details from the source

Reason for recall: Operative components of surgical systems potentially distributed with E-brake retainer pin, so in event of a structural column failure, the E-brake pin would prevent E-brake engagement and may result in uncontrolled motion. In addition, if it moves/backs out it may cause restricted column range of motion and/or uncontrolled motion, which may cause patient tissue injury/crushing injury/bleeding. Distribution: Worldwide - US Nationwide distribution including in the states of CA, GA, OH, PA, NY, MN, TN, OR, IL, MD, NJ, AL, NV, TX, FL, LA, CT, NC, AZ, IA, MO, WA, MI, WV, WI, KY, KS, VA, OK, MA, IN, CO, SC, SD, MS, NH, UT, ME and the countries of Belgium, Denmark, Sweden, France, Japan, Germany, Israel, Switzerland, South Korea, Czech Republic, Austria, United Kingdom, Italy, Spain, Ireland, Saudi Arabia, Quantity: 454 Lot/code info: UDI: 00886874114605. Serial Numbers: 605408, 606848, 612328, 628399, 609353, 614073, 617794, 616769, 10008062, 635659, 631694, 638771, 641430, 694998, 645433, 643209, 954097, 656203, 764644, 656319, 946098, 667190, 666480, 655562, 664662, 680530, 679266, 683974, 674171, 676962, 682694, 686777, 701248, 696877, 688725, 688225, 697922, 700316, 698277, 702699, 709536, 701751, 717804, 705366, 704731, 724178, 703152, 722803, 693262, 736468, 736338, 740359, 742366, 732233, 696089, 724835, 716520, 709987, 728391, 751529, 731765, 672743, 766366, 711517, 735003, 735252, 746457, 746623, 718940, 721722, 691190, 744094, 750174, 738718, 10211920, 759253, 761448, 752536, 812651, 754935, 790597, 811147, 793019, 761000, 812650, 919988, 919987, 913011, 913012, 926423, 759791, 926422, 938794, 771653, 938795, 766365, 963841, 797872, 769736, 794437, 970568, 970562, 977045, 10003059, 957781, 973824, 804588, 10009706, 936269, 10003509, 930240, 936268, 800142, 10003510, 982782, 967667, 10009707, 973355, 944585, 10065786, 935214, 10020729, 10012963, 10020705, 10031145, 10031144, 973825, 10012962, 10036960, 963856, 10711066, 10035476, 795038, 10043635, 944583, 10072852, 10072853, 966079, 10069899, 10076490, 10087446, 10054478, 10080929, 954096, 10130089, 10165287, 942811, 10180162, 10172589, 10103483, 10103484, 10056518, 10092007, 10246370, 10209961, 10183740, 10232634, 504007, 10231116, 10211880, 10194911, 10245457, 10112470, 10189443, 977044, 10238314, 10644376, 10036961, 10274101, 10112471, 1023831

UPC codes

00886874114605

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify