FDA (Device enforcement)Recalled 2026-05-21class ii

BD SWITZERLAND SARL: BD Connecta" BD Luer-Lok" 360 REF 394910 UDI-DI code: 00382903949106 BD Connecta" Stopcocks are used to administer fluids or medication through one or two different ports via an I

Hazard

Due to stopcock issues when using the plastic luer-lok syringe, the connection at the Port C may over thread or continue to spin rather than tightening securely. This behavior has led some clinical users to perceive the connection as unstable.

Details from the source

Reason for recall: Due to stopcock issues when using the plastic luer-lok syringe, the connection at the Port C may over thread or continue to spin rather than tightening securely. This behavior has led some clinical users to perceive the connection as unstable. Distribution: U.S.: AK, AL, AR, AZ, CA, CO, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV. O.U.S.: Japan Quantity: 2,465,500 units Lot/code info: Catalog/Model Number: 394910 UDI-DI code: 00382903949106 Lot Numbers: 4277968 4277973 4282451 4282461 4289801 4290582 4292904 4295848 4299949 4303841 4305108 4306700 5129635 5129642 5129649 5129650 5129651 5139250 5139251 5139252 5150181 5153940 5164061 5164062 5178078 5181442 5183810 5183816 5183824 5183828 5218287 5218291 5218293 5218299 5218302 5241476 5247729 5247731 5251098 5275601 5275602

UPC codes

00382903949106

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify