← Search recalls
FDA (Device enforcement)Recalled 2024-06-28class ii

Baxter Healthcare Corporation: Baxter PST 500 U, Precision Surgical Table, Part number (product code) 4080300

Hazard

Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the PST 500 U Precision Surgical Table due to a potential issue with loose spring pins in the spindle drives. This can cause the tabletop to unexpectedly tilt or move at any time, even without active use of the surgical table. This could potentially result in unintentional patient movement during surgical procedures and/or preparation for surgical procedures, including transport.

Details from the source

Reason for recall: Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the PST 500 U Precision Surgical Table due to a potential issue with loose spring pins in the spindle drives. This can cause the tabletop to unexpectedly tilt or move at any time, even without active use of the surgical table. This could potentially result in unintentional patient movement during surgical procedures and/or preparation for surgical procedures, including transport. Distribution: Worldwide distribution - US Nationwide and the country of Canada. Quantity: 400 units Lot/code info: UDI/DI 00887761973671, Serial Numbers: 107936298, 107936299, 107936300, 107946575, 107946576, 107949857, 107951075, 107952514, 107954982, 108021169, 108023287, 108028575, 108028716, 108028747, 108028798, 108028801, 108028803, 108028817, 108028829, 108028840, 108028846, 108029046, 108029052, 108029053, 108049023, 108049024, 108049948, 108049950, 108049955, 108049967, 108050823, 108050824, 108055575, 108055576, 108055577, 108055581, 108123400, 108127380, 108129826, 108129828, 108131719, 108131745, 108133272, 108133273, 108160433, 108160447, 108160448, 108160452, 108160453, 108160454, 108160455, 108167451, 108167552, 108167553, 108167554, 108167561, 108167597, 108167598, 108167701, 108167702, 108167706, 108167710, 108167715, 108167719, 108167723, 108260659, 108260711, 108260712, 108260713, 108260749, 108260750, 108260756, 108260757, 108260768, 108260769, 108260770, 108260771, 108260772, 108260910, 108260912, 108260913, 108260914, 108260915, 108260922, 108260925, 108260926, 108260927, 108260929, 108260933, 108260934, 108260935, 108260936, 108260937, 108294079, 108294080, 108294081, 108294082, 108294083, 108294084, 108294085, 108296714, 108296720, 108296734, 108296735, 108296736, 108296737, 108296738, 108296739, 108296740, 108296741, 108296742, 108296743, 108296744, 108296745, 108310315, 108310316, 108310317, 108310318, 108310320, 108310321, 108310322, 108339152, 108339153, 108339154, 108339155, 108339156, 108339157, 108339158, 108352860, 108352861, 108352866, 108352870, 108352871

UPC codes

00887761973671

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify