Visualase Cooled Laser Applicator System (VCLAS) labeled as: a. Part Number 9735559, Visualase¿ Cooled Laser Applicator System 3mm Tip; b. Part Number 9735560 Visualase¿ Cooled Laser Applicator System 10mm Tip; c. Part Number 9735561 Visualase¿ Cooled Laser Applicator System 15mm Tip Product Usage:
Check against the official notice below — it lists the full affected range.
Medtronic is recalling certain lots of Visualase Cooled Laser Applicator System (VCLAS) laser diffusing fibers in 3mm, 10mm, and 15mm sizes due to a calibration issue with test equipment. Because these fibers were not tested at the intended laser power levels, they may produce low laser energy output or experience fiber burnout when operated at lower wattage settings. A total of 1,283 devices have been distributed worldwide, including across most U.S. states and to Israel.
The affected devices are identified by specific part numbers (9735559 and 9735560), UDI codes, and lot codes listed in the recall notice. The source document does not specify what remedy affected consumers should take.
Written by software from the official notice below. Not reviewed by a lawyer.
Medtronic has become aware that certain lots of the VCLAS 3MM, 10MM, and 15MM laser diffusing fibers were not tested at intended laser power levels, due to an issue related to calibration of the test equipment, which can lead to the potential for low laser energy output from fiber or fiber burnout at lower wattage.
Reason for recall: Medtronic has become aware that certain lots of the VCLAS 3MM, 10MM, and 15MM laser diffusing fibers were not tested at intended laser power levels, due to an issue related to calibration of the test equipment, which can lead to the potential for low laser energy output from fiber or fiber burnout at lower wattage. Distribution: Worldwide Distribution - US Nationwide in the states of AL, AR, AZ, CA, CO, DC, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, NC, NJ, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI. International distribution to Israel. Quantity: 1,283 devices total Lot/code info: a. Part Number 9735559, UDI (00643169611986, 00643169933361), Lot Codes (0212782059, 0212894256, 0213251294, 0213331774, 0213383266, 0212783784, 0212926008, 0213259125, 0213335148, 0213383267, 0212823883, 0212956163, 0213299882, 0213375479, 0213387215, 0212823919, 0212978141, 0213299885, 0213378982, 0213537710, 0212854585, 0212996201, 0213327285, 0213383244, 0213966003, 0212882619, 0213519510); b. Part Number 9735560, UDI (00643169611993, 00643169933385), Lot Codes (0212086315, 0213158249, 0213299883, 0213427083, 0213728326, 0212698642, 0213158631, 0213304570, 0213455321, 0213735817, 0212715926, 0213158719, 0213304571, 0213458938, 0213737565, 0212745633, 0213165510, 0213304573, 0213507022, 0213760863, 0212753561, 0213165511, 0213308717, 0213507023, 0213760864, 0212756095, 0213205023, 0213309087, 0213512428, 0213760866, 0212760821, 0213205024, 0213334598, 0213512430, 0213769912, 0212827856, 0213208868, 0213368255, 0213513054, 0213917334, 0212838837, 0213212692, 0213368499, 0213513055, 0213922188, 0212843810, 0213223891, 0213368573, 0213513056, 0213924407, 0212846727, 0213223892, 0213373657, 0213540031, 0213924418, 0212856596, 0213242839, 0213374559, 0213576515, 0213928892, 0212873863, 0213245723, 0213375535, 0213576516, 0213945813, 0212873873, 0213258625, 0213375537, 0213576517, 0213945814, 0212879480, 0213259039, 0213375538, 0213594271, 0213966370, 0212913269, 0213259673, 0213375991, 0213594272, 0213972422, 0212922387, 0213259674, 0213376135, 0213594273, 0213973527, 021293527