FDA (Device enforcement)Recalled 2026-05-04class ii

Medline Industries, LP: MEDLINE Medical Procedure Kits labeled as: 1) KIT UROLOGY URETHROPLASTY, Medline Kit Number/SKU DYKMBNDL117C; 2) URETHROPLASTY, Medline Kit Number/SKU DYNJ908978; 3) URETHROPLA

Hazard

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

Details from the source

Reason for recall: During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification. Distribution: Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI. Quantity: 91 units Lot/code info: Medline Kit Number/SKU DYKMBNDL117C: UDI/DI 10195327371104 (each), 40195327371105 (case), Lot Number 23DLA834; Medline Kit Number/SKU DYNJ908978: UDI/DI 10195327231132 (each), 40195327231133 (case), Lot Number 23BMC463; Medline Kit Number/SKU DYNJ908978: UDI/DI 10195327231132 (each), 40195327231133 (case), Lot Number 23AMA685; Medline Kit Number/SKU DYNJ908978: UDI/DI 10195327231132 (each), 40195327231133 (case), Lot Number 23AMB905; Medline Kit Number/SKU DYNJ908978: UDI/DI 10195327231132 (each), 40195327231133 (case), Lot Number 22LMG466; Medline Kit Number/SKU DYNJ908978: UDI/DI 10195327231132 (each), 40195327231133 (case), Lot Number 22LMB983; Medline Kit Number/SKU DYNJ908978F: UDI/DI 10198459228803 (each), 40198459228804 (case), Lot Number 25JBD225; Medline Kit Number/SKU DYNJ908978F: UDI/DI 10198459228803 (each), 40198459228804 (case), Lot Number 25HBW236.

UPC codes

10195327371104, 40195327371105, 10195327231132, 40195327231133, 10198459228803, 40198459228804

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify