← Search recalls
FDA (Device enforcement)Recalled 2026-05-04class ii

Medline Industries, LP: MEDLINE Medical Procedure Kits labeled as: BILAT MICRO VASECTOMY REVERSAL, Medline Kit Number/SKU DYNJ68302

Hazard

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

Details from the source

Reason for recall: During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification. Distribution: Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI. Quantity: 588 units Lot/code info: UDI/DI 10193489820478 (each), 40193489820479 (case), Lot Numbers: 24CBL413, 24ABX694, 23JBY172, 23IBI716, 23DBG873, 23BBR326, 23ABH376, 22KBS634, 22JBS743, 22FBT429, 22BBZ842, 22BBK836, 21BBG579, 21ABO658.

UPC codes

10193489820478, 40193489820479

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify