Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
Reason for recall: Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral. Distribution: US Nationwide distribution in the state of TX. Quantity: 3 units Lot/code info: Model No. P002-62; UDI-DI: 05407008320041; Serial No. 1000 to 1690.