← Search recalls
FDA (Device enforcement)Recalled 2020-05-15closedclass ii

Siemens Healthcare Diagnostics, Inc.: (1) Atellica Data Manager software version 1.0 Siemens Material Number (SMN): 11314237 (2) Atellica Data Manager software version 1.1 Siemens Material Number (SMN

Is my product affected?

(1) Atellica Data Manager software version 1.0 Siemens Material Number (SMN): 11314237 (2) Atellica Data Manager software version 1.1 Siemens Material Number (SMN): 11316888 Product Usage: are clinical laboratory information modules which consolidate test orders and test results, provide test orderi

  • UPC 00630414234946
  • UPC 006304142790

Check against the official notice below — it lists the full affected range.

What happened

Siemens Healthcare Diagnostics is recalling Atellica Data Manager software versions 1.0 and 1.1 due to an unexpected interface driver behavior that may cause quality control (QC) results to be assigned to an incorrect control lot number. This defect could result in the reporting of erroneous patient results if a QC test failed but appeared to be passing and the issue went undetected during QC review. The software has been distributed worldwide to 70 units across numerous countries including the United States, Argentina, Australia, Belgium, Brazil, Chile, Colombia, and many others.

The recall document does not specify a remedy or recommended actions for affected consumers.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

Unexpected interface driver behavior identified (QC) Results May Be Assigned to an Incorrect Control Lot Number, and lead to the reporting of erroneous patient results if the QC failed but appeared to be passing and the issue is not detected during QC review

Details from the source

Reason for recall: Unexpected interface driver behavior identified (QC) Results May Be Assigned to an Incorrect Control Lot Number, and lead to the reporting of erroneous patient results if the QC failed but appeared to be passing and the issue is not detected during QC review Distribution: Worldwide distribution - US Nationwide distribution and the countries of Argentina, Australia, Bahamas, Belgium, Brazil, Chile, Colombia, Czech Republic, Egypt, Germany, Greece, Hungary, India, Italy, Kuwait, Latvia, Lithuania, Malaysia, Mexico, Oman, Peru, Poland, Portugal, Qatar, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Thailand, Turkey, United Arab Emirates, United Kingdom. Quantity: 70 units Lot/code info: Software versions: Atellica¿ Data Manager software version 1.0 UDI: 00630414234946 Atellica¿ Data Manager software version 1.1 UDI: 006304142790

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.