Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.
Reason for recall: Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps. Distribution: Domestic distribution nationwide. International distribution to Canada. Quantity: 930 Lot/code info: Lot Code: Model No 8713040U; UDI-DI 4046963687182; All Serial Numbers
4046963687182
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.