Users were reporting instances of "Gas Loss in IAB Circuit" and "Gas Gain in IAB Circuit" alarms while providing therapy. An internal investigation of the complaints determined that there is a potential trigger for these alarms that was not listed in the IFU: Patient Movement (coughing, general movement, and swallowing). Blood in the circuit, kinks/abrasions of the tubing/catheter and certain patient conditions could also trigger these alarms. The firm's HHE also identified hardware issues related to these alarms.
Reason for recall: Users were reporting instances of "Gas Loss in IAB Circuit" and "Gas Gain in IAB Circuit" alarms while providing therapy. An internal investigation of the complaints determined that there is a potential trigger for these alarms that was not listed in the IFU: Patient Movement (coughing, general movement, and swallowing). Blood in the circuit, kinks/abrasions of the tubing/catheter and certain patient conditions could also trigger these alarms. The firm's HHE also identified hardware issues related to these alarms. Distribution: Worldwide - US Nationwide distribution. Quantity: 9175 units Lot/code info: All Lot Numbers. Model: 0998-00-0800-75, UDI: 10607567112312; Model: 0998-00-0800-83, UDI: 10607567108407; Model: 0998-00-0800-85, UDI: 10607567113449