FDA (Device enforcement)Recalled 2025-07-14class ii

MAQUET GMBH: Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2; All Software Versions.

Hazard

Under certain conditions, such as the simultaneous use of the operating table with other medical devices, including high-frequency (HF) surgical equipment, interference may occur that could prevent the device from performing as intended. If the issue occurs, the operating table is unresponsive to command from both the universal remote control and override panel.

Details from the source

Reason for recall: Under certain conditions, such as the simultaneous use of the operating table with other medical devices, including high-frequency (HF) surgical equipment, interference may occur that could prevent the device from performing as intended. If the issue occurs, the operating table is unresponsive to command from both the universal remote control and override panel. Distribution: Worldwide - US Nationwide distribution in the states of KY, NY, TX and the countries of Algeria, Australia, Austria, Bulgaria, Canada, China, Colombia, Denmark, Egypt, Estonia, Finland, France, Germany, Ireland, Israel, Italy, Japan, Jordan, Korea, Republic of, Kuwait, Lebanon, Morocco, Netherlands, Norway, Poland, Portugal, Romania, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerlan Quantity: 315 units (8 US, 307 OUS) Lot/code info: Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2; UDI-DI: (1) 04046768145177, (2) 04046768145184, (3) 04046768145191, (4) 04046768145207, (5) 04046768145214, (6) 04046768145221; Lot/Serial Numbers: (1) 1, 2, 3, 10, 25, 26, 27, 28, 22, 4, 5, 6, 7, 8, 9, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 24, (2) 9, 51, 53, 14, 15, 16, 17, 34, 35, 36, 37, 39, 40, 42, 43, 54, 19, 20, 21, 22, 23, 24, 25, 26, 27, 28, 29, 30, 31, 32, 33, 38, 44, 45, 46, 5, 13, 57, 58, 59, 18, 49, 47, 48, 7, 52, (3) 41, 5, 6, 45, 48, 49, 50, 23, 38, 10, 12, 13, 14, 84, 87, 88, 89, 7, 63, 80, 82, 83, 86, 54, 93, 94, 11, 15, 18, 20, 21, 22, 27, 28, 32, 40, 9, 69, 70, 71, 72, 73, 74, 75, 76, 77, 78, 79, 85, 55, 56, 57, 58, 59, 42, 97, 92, 17, 26, 16, 24, 25, 30, 33, 35, 36, 37, 39, 44, 53, 91, 8, 19, 90, 46, 47, 51, 52, 34, 43, 60, 62, 64, 65, 66, 67, 68, (4) 44, 70, 71, 42, 56, 65, 115, 185, 52, 105, 106, 107, 109, 145, 146, 147, 148, 11, 26, 19, 32, 33, 35, 36, 39, 53, 61, 63, 69, 14, 20, 68, 13, 18, 16, 101, 167, 170, 83, 3, 4, 23, 28, 37, 38, 40, 41, 43, 45, 47, 48, 50, 55, 58, 60, 62, 78, 81, 82, 92, 94, 96, 97, 98, 99, 100, 102, 116, 138, 139, 144, 150, 152, 153, 154, 155, 156, 165, 169, 171, 21, 22, 24, 25, 6, 17, 27, 29, 30, 34, 49, 51, 57, 64, 74, 75, 76, 77, 79, 80, 84, 85, 86, 87, 88, 89, 90, 91, 93, 95, 103, 104, 108, 110, 111, 112, 113, 114, 118, 119, 120, 121, 122, 123, 124, 125, 126, 127, 128, 129, 130, 131, 132, 133, 134, 135, 136, 137, 140, 141,

UPC codes

04046768145177, 04046768145184, 04046768145191, 04046768145207, 04046768145214, 04046768145221

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify